Table 6.

Peak signal-intensity (C0 + pre-contrast value) plus or minus one standard deviation for the experiment in arbitrary units. For α = 0.05, tα/2 = 2.776, and |t| > tα/2 implies statistically that the means reported by the two methods are significantly different (highlighted in bold). Key: PA - Pulmonary Artery, LL - Left Lung, RL - Right Lung, PV - Pulmonary Vein, DA - Descending Aorta

Region Sliding
Window
Proposed Mean
Diff.
Std. Dev.
Diff.
t test
statistic
PA 57.67 ±12.27 s 59.37 ±12.08 s 1.6998 s 0.63395 s 5.9955
LL 33.29 ±10.88 s 33.68 ±11.34 s 0.39262 s 0.70342 s 1.2481
RL 35.50 ±6.99 s 35.94 ±7.30 s 0.43356 s 0.5662 s 1.7122
PV 75.18 ±12.73 s 74.13 ±12.77 s −1.0464 s 0.8097 s 2.8898
DA 78.03 ±20.53 s 76.73 ±21.45 s −1.2922 s 2.8812 s −1.0029