Published in final edited form as: Talanta. 2010 Aug 13;82(5):1892–1904. doi: 10.1016/j.talanta.2010.08.005

Table 2.

The concentrations of R,S)-Ket, (R,S)-norKet, (R,S)-DHNK, (2R,6R)-/(2S,6S)-HNK, (2S,6S)-HKet and (2S,6R)-HKet in the calibration standards and quality control standards used in the validation of the achiral method in plasma.

(R,S)-Ket (ng/ml) (R,S)-norKet (ng/ml) (R,S)-DHNK (ng/ml) (2R,6R)-/(2S,6S)-HNK (ng/ml) (2S,6S)-HKet (ng/ml) (2S,6R)-HKet (ng/ml)
Calibration
1 25,000 500 500 2,000 40 30
2 18,750 375 375 1,500 30 22.5
3 12,500 350 350 1,000 20 15
4 6,250 125 125 500 10 7.5
5 5,000 94 94 375 7.5 5.6
6 3,125 63 63 250 5 3.8
QC
High 21,875 437 437 1,750 35 26
Medium 15,625 312 312 1,250 25 18
Low 9,375 187 187 750 15 11