TABLE IV.

Measured Acoustic Tank Performance for Control and Low-Loss 3-Layer Ceramic Phased Arrays

PZT type PZT-4 PZT5-H
Backing Z (MRayl) 0.7 3.2
Sensitivity (dB re input) −63.7 −67.5
Fc (MHz) 3.3 3.4
BW (%) 60 66
t-6 dB (μs) 0.47 0.37
t-20 dB (μs) 1.34 1.24
t-30 dB (μs) 2.41 1.80
t-40 dB (μs) 3.61 3.38