Skip to main content
. 2010 Oct 1;5(10):e13125. doi: 10.1371/journal.pone.0013125

Table 1. Comparison of µFBI and standard bead-based assay.

Procedure Micro fluidic bead-based assay Standard bead-based assay Fold reduction
Time Usage per capillary Time Usage per well
Pre-wet 10 min 20 µL 5 min 100 µL
Beads 1 µL, 1000beads/analyte 30 µL, 2000beads/analyte
Vacuum - -
Sample 60 min 1 µL, 200 ug/ml 60 min 100 µL, 200 ug/ml 100×
3×wash - -
Detection antibody 60 min 1 µL, 1× 60 min 30 µL, 1× 30×
3×wash - -
Strep-PE 60 min 1 µL, 1× 60 min 30 µL, 1× 30×
3×wash - -
Recover beads 3 min 5 µL - -
Measurement 80 µL, 80 s, Gate: 7500–15000 80 µL, 80 s, Gate: 7500–15000

-: without this procedure.