Published in final edited form as: Int J Radiat Oncol Biol Phys. 2010 Jul 7;79(1):312–320. doi: 10.1016/j.ijrobp.2010.03.011

Table 1.

Summary of the average number of values failing the 2% dose difference 2 mm distance to agreement γ-test for all experiments

Plan type and motion trace With tracking Without tracking % reduction p-value
Low modulation lung 4.1% 59% 93% <0.001
High modulation lung 9.1% 56% 84% <0.001
   Combined lung 6.6% 58% 89% <0.001
Low modulation prostate 0.6% 31% 98% <0.001
High modulation prostate 11% 36% 70% <0.001
  Combined prostate 5.7% 34% 83% <0.001
   All combined 6.1% 46% 86% p<0.001