Table 2.
Summary of the average number of values failing the 3% dose difference 3 mm distance to agreement γ-test for all experiments
| Plan type and motion trace | With tracking | Without tracking | % reduction | p-value |
|---|---|---|---|---|
| Low modulation lung | 0% | 34% | 100% | <0.001 |
| High modulation lung | 3.2% | 35% | 91% | <0.001 |
| Combined lung | 1.6% | 34% | 95% | <0.001 |
| Low modulation prostate | 0% | 12% | 100% | 0.009 |
| High modulation prostate | 2.4% | 16% | 85% | <0.001 |
| Combined prostate | 1.2% | 14% | 93% | <0.001 |
| All combined | 1.4% | 24% | 94% | p<0.001 |