Published in final edited form as: AJR Am J Roentgenol. 2009 Nov;193(5):1340–1345. doi: 10.2214/AJR.09.2886

TABLE 1. Results of Simulated Scans.

Breast Dose (mGy)a Lung Dose (mGy)

Patient Perimeter (cm) Patient Age (y) Fixed Tube Current Tube Current Modulation Percentage Dose Reduction Fixed Tube Current Tube Current Modulation Percentage Dose Reduction

81 63 29 13 57 34 17 49
87 57 23 8 64 27 12 56
88 35 22 9 60 26 12 53
89 64 21 12 41 26 17 36
89 67 26 13 50 31 16 47
92 41 24 14 43 31 19 39
92 16 19 9 52 24 13 44
93 89 22 8 62 25 12 52
98 71 16 14 11 22 19 14
99 61 21 15 27 25 18 27
100 52 19 13 34 24 14 42
101 28 22 19 14 24 19 22
101 78 20 12 40 26 18 30
112 62 20 18 10 25 25 2
114 75 16 14 8 17 18 −2
114 70 18 12 34 25 16 39
115 35 16 11 30 22 17 20
116 77 18 15 15 22 19 15
119 71 16 18 −12 22 22 0
121 44 16 15 5 18 18 0
122 56 17 15 10 21 19 8
123 49 16 18 −12 19 21 −10
124 31 16 19 −15 20 22 −8
125 70 19 19 0 22 23 −2
128 39 14 20 −41 16 22 −33
128 29 16 18 −14 18 20 −14
132 69 15 16 −8 17 18 −2
134 62 14 17 −21 19 22 −16
135 55 14 16 −15 17 19 −12
142 62 14 16 −16 14 16 −16

111 (mean) 56 19 15 17 23 18 16
81 (minimum) 16 14 8 -41 14 12 -33
142 (maximum) 89 29 20 64 34 25 56
a

Breast dose estimates from [27].