Box 1.

Various settings for VEP

(In full compliance with the recommendations of the International Federation of Clinical Neurophysiology (IFCN) Committee[14])

Recording conditions:
1. Filter:
  1. Low cutoff (high pass): 1–3 Hz

  2. High cut off (low pass): 100–300 Hz

2. Amplification between 20,000 and 1,00,000
3. Sweep duration between: 250 and 500 ms
4. Number of epochs: at least 100 were averaged
5. Electrode impedance was kept below 5 kohms.
Stimulation options:
1. Black and white checker board
2. Distance between subject and screen: 70–100 cm
3. Contrast: 50%–80%
4. Fixation point for full field size greater than 8°
5. Size of pattern element: 8 × 8 mm.
6. Rate of stimulation:
  1. For transient VEP: 1 Hz

  2. For steady state VEP: 4–8 Hz

7. Mean luminance of the central field: 20–40 cd/m2
Background luminance: 20–40 cd/m2
Two channels (as per 10–20 international system)
1. Channel 1: Oz–FPz
2. Channel 2: Oz–A1A2 (linked ear)
Ground electrode position at: Cz