TABLE 6.

Key performance characteristics of the NCEH's LC-MS/MS measurement procedure as of July 20091

25(OH)D3 25(OH)D2
Accuracy
 Compared with SRM 9722 +2 to + 83 −18 to 303
 Recovery (mean ± SD) 95 ± 3% 97 ± 3%
Precision,4% 6–85 6–96
Sensitivity, nmol/L <2.5 <2.5
Specificity Yes7 Yes
1

The LC-MS/MS measurement procedure is in development. It has not yet been used to measure serum 25(OH)D in NHANES participants. When development is completed, the LC-MS/MS will be used to measure serum 25(OH)D in NHANES 2007 and later.

2

SRM 972 certificate values (15) compared with CDC data (12 replicates per level over 3 d) from assays performed in August to September 2008.

3

Values expressed as percent of SRM 972.

4

NCEH coefficient of variation data derived from 29 assays performed in September 2008 to July 2009.

5

The 25(OH)D3 concentrations ranged from 31 to 85 nmol/L. Unit conversion: 2.5 nmol/L = 1 ng/mL.

6

The 25(OH)D2 concentrations ranged from 2 to 22 nmol/L; precision for 7–22 nmol/L. Unit conversion: 2.5 nmol/L = 1 ng/mL.

7

Measurement procedure does not separate serum 25(OH)D3 from 3-epi-25(OH)D3.