Skip to main content
. 2008 Aug;6(4):557–567. doi: 10.1089/adt.2008.139

Table 1.

Dynamic Range of Detected Signal as Fold Change (x)

 
Platform
 
IN Cell 1000
DTX
EnVision
Dye Signal Cell titer Signal Cell titer Signal Cell titer
DAPI 135.3x 32x 53.6x 32x 83.0x 32x
eGFP 23.3x 32x 2.2x 4x 3.9x 16x
DsRED 66.6x 32x 2.9x 4x 7.1x 16x

Dynamic range represents the quotient of instrument maximum and minimum values obtained from cell titers residing on a positive trendline and with P < 0.001.