Table 2.

Comparison of Point and Trend Accuracy Parameters for Different Commercially Available CGM Devices

Point accuracy – Euglycemia, 70 < BG < 180 mg/dl (3.9–10.0 mmol/liter)a
Guardian®b DexCom STS®b DexCom SEVEN PLUS®c Navigator®b GlucoDay®Sb GlucoMen®Day
Mean absolute error mg/dl 16.4 22.3 n.a.d 16.0 15.7 11.9
mmol/liter 0.91 1.24 n.a. 0.89 0.87 0.66
Mean absolute relative error (%) 15.2 21.2 15.0 15.3 15.6 10.4
Median absolute error mg/dl 14.8 19.1 n.a. 15.3 10.8 10.1
mmol/liter 0.82 1.06 n.a. 0.85 0.60 0.56
Median absolute relative error (%) 13.3 18.4 13.0 11.8 10.7 7.7
% readings within ISO 15197 requirementse 73.2 52.2 74.0 72.2 76.9 89.3
Point Accuracy – Hypoglycemia, BG ≥ 70 mg/dl [3.9 mmol/liter]f
Mean absolute error mg/dl 9.9 13.1 n.a. 6.5 8.5 7.4
mmol/liter 0.55 0.73 n.a. 0.36 0.47 0.41
Mean absolute relative error (%) 16.1 21.5 25.0 10.3 17.5 14.2
Median absolute rrror mg/dl 7.6 11.52 n.a. 4.3 7.2 5.9
mmol/liter 0.42 0.64 n.a. 0.24 0.40 0.33
Median absolute relative error (%) 13.8 22.5 20.0 7.4 15.6 9.6
% readings within ISO 15197 requirementse 76.5 52.9 n.a. 79.4 83.0 80.0
Rate accuracy – Descent into hypoglycemia
Absolute rate deviation mg/dl/min 0.87 0.72 n.a. 0.66 1.74 0.75
mmol/liter/h 2.9 2.4 n.a. 2.2 5.8 2.5
Rate accuracy – Ascent from hypoglycemia
Absolute rate deviation mg/dl/min 0.90 0.99 n.a. 0.99 2.79 0.45
mmol/liter/h 3.0 3.3 n.a. 3.3 9.3 1.5
a

DexCom SEVEN PLUS system assumes the euglycemic range to be 80–180 mg/dl (4.4–10.0 mmol/liter).30

b

Data from Kovatchev BP, Anderson S, Heinemann L, Clarke WL. Comparison of the numerical and clinical accuracy of four continuous glucose monitors. Diabetes Care. 2008;31:1160–4.20

c

Data from: SEVEN PLUS User Guide. Dexcom, Inc: San Diego, CA.30

d

n.a., data not available

e

Overall percentage of readings falling within ±15 mg/dl (if <75 mg/dl) and ±20% (if ≥75 mg/dl) of the reference values, according to the ISO 15197 accuracy requirements (bias plot).23 ISO, International Organization for Standardization.

f

DexCom SEVEN plus system defines as hypoglycemia BG levels ≥80 mg/dl (≥4.4 mmo/liter).30