Skip to main content
. 2010 Sep 1;4(5):1265–1268. doi: 10.1177/193229681000400528

Table 1.

Accuracy of Common Monitoring Systems in the United States

System Percentage meeting standard (above 75 mg/dl; absolute value in parenthesis below 75 mg/dl)
20% (15 mg/dl) 15% (12 mg/dl) 10% (8 mg/dl) 5% (4 mg/dl)
A 99 91 68 35
B 96 91 82 52
C 99 95 84 53
D 100 98 88 52
E 99 98 92 63
F 98 97 89 61
G 98 95 79 41
H 99 98 92 65
I 99 98 93 71