Table 2. MLPA test for CLL: interpretative criteria and calling guidelines.
Normalized probe ratio | Calling guidelines | Interpretation, p-value |
Within Mean ± 2SD | Duplicates show > = 50% probes in the cluster are within this range | Normal |
Out of Mean ± 3SD | Duplicates show > = 1 probe in the cluster are out of this range | Deletion or Amplification, P<0.01 |
Between Mean ± 2SD and Mean ± 3SD | Duplicates show > = 50% probes in the cluster are within this range | Loss or Gain, P<0.05 |