TABLE I. ASSESSMENT OF THE CONDITION OF CUFFS AND BLADDERS.
Component | Test |
Satisfactory n = 10 (%) |
Unsatisfactory n = 10 (%) |
---|---|---|---|
Cuff | Examine general condition of the cuff and ability to fasten around arm | 8 (80) | 2 (20) |
Bladder size | Measure bladder dimensions and compare with standard | 0 (0) | 10 (100) |
State of rubber components | Check the state of the rubber components | 5 (50) | 5 (50) |
Pump bulb – operation of inlet valve | Check operation of inlet valve | 1 (10) | 9 (90) |
Control valve operation | Check operation of control valve | 5 (50) | 5 (50) |