Table 1.
Sequence: | Single-shot gradient-echo echoplanar imaging |
Image acquisition parameters | TR ≤ 1500 ms TE = 35 to 45 ms at 1.5T TE = 25 to 30 ms at 3T Flip angle = 60° to 90° at 1.5T Flip angle = 60° at 3T |
Image acquisition duration | 90 to 120 seconds Image acquisition started 10 seconds before initiation of bolus injection to achieve at least 10 baseline images |
Coverage and slice thickness | Field of view ≈ 24 cm Whole brain coverage using ≥ 12 slices Gap 0 to 1 mm Slice thickness 5 mm Matrix size 128 × 128 |
Slice orientation | Parallel to hard palate |
Contrast material | Standard Gadolinium-based contrast material |
Contrast volume | Single dose (for half molar agent ≈ 20 mL for 100 kg person) followed by 20 to 40 mL saline flush |
Injection rate | 4 to 6 mL per second. Same rate for saline |
IV access | 18 to 20 gauge IV line. Right antecubital vein preferred |