TABLE 1.

Absorbed Fraction of Uniformly Distributed 131I Source in Various Sizes of Spheres Filled with Water as Medium

Sphere R (cm) 131I β* Voxel sphere* Ref. A Sphere R (cm) 131 Voxel sphere Ref. B§
0.1 0.71 0.73 1.0 0.025 0.028 0.025
0.2 0.85 0.87 N/A 2.0 0.049 0.048 0.049
0.25 0.88 0.90 0.89 4.0 0.099 0.101 0.099
0.5 0.94 0.96 0.94 6.0 0.149 0.149 0.151
1.0 0.97 0.98 0.97 8.0 0.197 0.197 0.197
2.0 0.99 0.98 0.98 10.0 0.244 0.244 0.244
12.0 0.287 0.288 0.288
14.0 0.329 0.328 0.329
*

NPS = 10,000; relative errors < 10%.

Akabani et al. (32).

NPS = 50,000; relative errors < 10%.

§

Furhang et al. (33).

Sphere with 0.1-cm radius is below resolution of (0.4 cm)3 voxel.

NPS = number of histories; N/A = not applicable.