Skip to main content
. 2010 Jun 8;11(3):1018–1025. doi: 10.1208/s12249-010-9464-2

Table III.

Intra-day, Inter-day Precision and Accuracy of the HPLC Method Used for the Determination of In Vitro Tianeptine Sodium Concentration

Concentration (μg/ml) Precision C.V.% Accuracy
Intra-day Inter-day Ctheoretical Cmeasured % RE
10 1.97596 1.39142 10 9.86 −1.42
20 6.86578 1.85418 20 19.49 −2.54
30 4.91650 2.14179 30 28.78 −4.08
40 3.88590 1.39157 40 39.71 −0.73
50 4.31644 2.73502 50 48.76 −2.48
60 4.48191 0.89821 60 60.66 1.09

RE relative error