Table IV.

Validation of HPLC Technique for the Analysis of Carbamaz epine in Dogs Plasma

Parameters Values
Limit of detection (LOD; μg/ml) 0.025
Limit of quantitation (LOQ; μg/ml) 0.25
Linearity
 Range (μg/ml) 0.25–4.0
 Regression equation y = 4.8795 × −0.041
 Correlation coefficient (r) 0.9978–0.9992
 Recovery range (%) 70.45–80.43
Precision
 Intra-day R.S.D. (%)
 With sample injected 3 times 2.58
 With samples prepared 3 times 3.71
 Inter-day R.S.D. (%)
 With sample injected 3 times 5.74
 With samples prepared 3 times 7.43