Table I. Parameters of magnetic bead handler protocol for biotin-PE immunoassay.

Step Plate content Volume Mixing parameters
Time Temperature Speed
μl
1) Transfer Magnetic bead array 50
2) Analyte capture Sample 100 30 min RT 750 rpm
3) Wash + transfer Wash buffer 100 20 s RT Slow
4) Wash + transfer Wash buffer 100 20 s RT Slow
5) Bead release to assay plate Assay buffer 100 20 s RT Slow