Table II. Parameters of magnetic bead handler protocol for sandwich immunoassay.

Step Plate content Volume Mixing parameters
Time Temperature Speed
μl
1) Transfer Magnetic bead array 50
2) Analyte capture Crude cell culture lysate 100 Overnight 8 °C 750 rpm
3) Wash + transfer Wash buffer 100 20 s RT Slow
4) Wash + transfer Wash buffer 100 20 s RT Slow
5) Detection antibody (5 cycles) Detection antibody mixture 100 10 min, pause
2 min RT Very slow
6) Wash + transfer Wash buffer 100 20 s RT Slow
7) Wash + transfer Wash buffer 100 20 s RT Slow
8) Detection reagent (4 cycles) SAPE solution 100 10 min, pause
2 min RT Very slow
9) Wash + transfer Wash buffer 100 20 s RT Slow
10) Wash + transfer Wash buffer 100 20 s RT Slow
11) Bead release to assay plate Assay buffer 100 20 s RT Slow