Table III. Ambient analyte theory: repetitive measurement of biotin-PE from same sample.

Shown are the results for five sequentially performed analyses of biotin-PE from the same sample from the same well. The data from the first measurement were used for a five-parametric logistic fit. Using the parameters of this fit, the raw assay data were transformed to analyte concentrations. After five assays, an analyte decrease of 20% was observed in total, suggesting a captured amount of less than 5% per assay. AU, arbitrary units.

Biotin-PE First measurement
Second measurement
Third measurement
Fourth measurement
Fifth measurement
Mean %CV Mean %CV Mean %CV Mean %CV Mean %CV
pm AU AU AU AU AU
Blank 8 8 8 16 8 8 8 13 8 13
1 82 2 82 3 80 1 72 3 70 2
5 315 2 309 1 293 1 271 1 253 3
20 1139 2 1158 3 1071 0 1015 1 981 3
80 3466 2 3580 1 3366 1 3211 2 3131 1
300 6181 1 6191 1 5996 0 5926 2 5895 1
pm %CV pm %CV pm %CV pm %CV pm %CV
1 1 2 1 3 1 1 1 3 1 2
5 5 2 5 1 5 1 4 1 4 3
20 19 2 20 3 18 0 17 1 17 3
80 78 2 83 1 75 1 69 2 67 1
300 312 1 314 1 276 0 265 2 260 1