Skip to main content
. 2008 Aug;23(4):399–409. doi: 10.1089/cbr.2007.0457

Table 1.

Resolution of Reconstructed microPET Imagesa

 
 
Full width at half maximum (mm)
Full width at tenth maximum (mm)
Reconstruction method Distance from scanner axis (mm) radial tangential radial tangential
18F
 OSEM 0 1.84 ± 0.02b   3.49 ± 0.05c  
  8 2.05 ± 0.06c 1.94 ± 0.04c 3.69 ± 0.08d 3.56 ± 0.07d
  16 2.10 ± 0.05c 1.91 ± 0.03c 3.89 ± 0.08d 3.60 ± 0.07d
 MAP 0 1.88 ± 0.01b   3.52 ± 0.02c  
  8 2.01 ± 0.02c 1.99 ± 0.02c 3.71 ± 0.02d 3.69 ± 0.02d
  16 1.89 ± 0.01c 1.92 ± 0.03c 3.53 ± 0.02d 3.59 ± 0.03d
 3DRP 0 2.47 ± 0.02b   4.59 ± 0.04c  
  8 2.36 ± 0.03c 2.59 ± 0.03c 4.37 ± 0.03d 4.76 ± 0.04d
  16 2.51 ± 0.02c 2.42 ± 0.02c 4.63 ± 0.03d 4.50 ± 0.03d
124I
 OSEM 0 3.00 ± 0.06b   6.07 ± 0.33c  
  8 2.45 ± 0.17c 2.33 ± 0.04c 5.22 ± 0.28d 5.30 ± 0.53d
  16 2.97 ± 0.10c 3.01 ± 0.03c 5.92 ± 0.17d 5.54 ± 0.35d
 MAP 0 2.62 ± 0.06b   5.16 ± 0.17c  
  8 2.37 ± 0.15c 2.25 ± 0.12c 5.08 ± 0.81d 5.40 ± 0.54d
  16 2.35 ± 0.05c 2.67 ± 0.12c 4.92 ± 0.07d 5.50 ± 0.18d
 3DRP 0 3.47 ± 0.06b   7.53 ± 0.55c  
  8 3.10 ± 0.14c 2.99 ± 0.16c 8.06 ± 0.96d 6.88 ± 0.39d
  16 3.60 ± 0.14c 3.52 ± 0.09c 6.98 ± 0.35d 7.26 ± 0.45d

PET, positron emission tomography; OESEM, ordered subsets expectation maximum; MAP, maximum a posteriori; 3DRP, three-dimensional filtered backprojection.

a

Data are mean values and corresponding standard errors.

b

Number of measurements = 20.

c

Number of measurements = 10.

d

Number of measurements = 5.