Table 3.

results of inaccuracy (bias) and precision measurements. Inaccuracy (bias) is expressed in %; precision experiments in % relative standard deviation (RSD). Bias was within 15% of the nominal value at all concentration levels. For the precision experiments, the RSD results were lower than 15% at all concentration levels.

Adenosine(ng/ml) 1000 500 125 31.3 15.6
Inaccuracy (bias)(%) 1.35 2.65 0.66 2.04 0.19
Within-day precision (%RSD) 3.13 4,47 6.82 9.60 5.93
Between–day precision(%RSD) 2.49 4.18 0* 0* 5.15
*

If the ‘mean square between groups’ < ‘mean square within groups’, the RSD is set ‘0’. The mean squares are obtained from the analysis of variance-table [27].