Clinical and radiographic findings

Alumina/alumina CoCr/HXLPE p
HHS
 Preoperative (/100) 45.6 ± 8.2 43.1 ± 8.2 0.06
 Postoperative (/100) 95.1 ± 3.9 93.8 ± 5.1 0.07
  Pain none or slight (%) 93 95 0.75
  Limp none or mild (%) 98 96 0.67
  Patients’ satisfaction (%) 98 97 1.00
  Good and excellent (%) 96 94 0.48
Radiographic findings
 Femoral component
  RLL zone 1 (%) 1 7 0.08
  RLL zone 7 (%) 1 4 0.27
  Subsidence 0 0 1.00
  Instability 0 1 0.29
  Pedestal 18 23 0.63
 Acetabular component
  RLL zone 1 (%) 2 3 0.93
  RLL zone 2 (%) 1 0 0.34
  RLL zone 3 (%) 1 3 0.51
  Migration 0 0 1.00

CoCr cobalt chrome, HXLPE highly cross-linked polyethylene, RLL radiolucent line