Table 2.
Dialyzer | Dialysate flow rate (mL/min) | Clearance (mL/min)a | KoA (mL/min)b | Ko (cm/min)b |
---|---|---|---|---|
Optiflux F160NR | 350 | 232 ± 3 | 720 ± 32 | 0.046 ± 0.002 |
500 | 252 ± 7 | 703 ± 38 | 0.045 ± 0.002 | |
800 | 281 ± 3 | 776 ± 23 | 0.050 ± 0.001 | |
Optiflux F200NR | 350 | 237 ± 2 | 749 ± 20 | 0.038 ± 0.001 |
500 | 270 ± 4 | 847 ± 33 | 0.043 ± 0.002 | |
800 | 282 ± 5 | 785 ± 35 | 0.040 ± 0.002 | |
Revaclear | 350 | 233 ± 4 | 718 ± 31 | 0.053 ± 0.002 |
500 | 255 ± 5 | 713 ± 33 | 0.052 ± 0.002 | |
800 | 276 ± 3 | 732 ± 24 | 0.054 ± 0.002 | |
Revaclear MAX | 350 | 240 ± 3 | 789 ± 29 | 0.046 ± 0.002 |
500 | 270 ± 3 | 852 ± 30 | 0.050 ± 0.002 | |
800 | 293 ± 3 | 884 ± 30 | 0.052 ± 0.002 |
Clearances based on rate of disappearance from blood.
Calculated using the blood water flow rate (QBWi).
Data are presented as mean ± SEM for n = 12 observations. Statistical differences are described in the text.