Urea

Dialyzer Dialysate flow rate (mL/min) Clearance (mL/min)a KoA (mL/min)b Ko (cm/min)b
Optiflux F160NR 350 232 ± 3 720 ± 32 0.046 ± 0.002
500 252 ± 7 703 ± 38 0.045 ± 0.002
800 281 ± 3 776 ± 23 0.050 ± 0.001
Optiflux F200NR 350 237 ± 2 749 ± 20 0.038 ± 0.001
500 270 ± 4 847 ± 33 0.043 ± 0.002
800 282 ± 5 785 ± 35 0.040 ± 0.002
Revaclear 350 233 ± 4 718 ± 31 0.053 ± 0.002
500 255 ± 5 713 ± 33 0.052 ± 0.002
800 276 ± 3 732 ± 24 0.054 ± 0.002
Revaclear MAX 350 240 ± 3 789 ± 29 0.046 ± 0.002
500 270 ± 3 852 ± 30 0.050 ± 0.002
800 293 ± 3 884 ± 30 0.052 ± 0.002
a

Clearances based on rate of disappearance from blood.

b

Calculated using the blood water flow rate (QBWi).

Data are presented as mean ± SEM for n = 12 observations. Statistical differences are described in the text.