Skip to main content
. 2010 Jun 13;25(12):3990–3995. doi: 10.1093/ndt/gfq326

Table 3.

Phosphate

Dialyzer Dialysate flow rate (mL/min) Clearance (mL/min)a KoA (mL/min)b Ko (cm/min)b
Optiflux F160NR 350 168 ± 6 434 ± 34 0.028 ± 0.002
500 178 ± 9 450 ± 62 0.029 ± 0.004
800 194 ± 8 458 ± 33 0.029 ± 0.002
Optiflux F200NR 350 178 ± 8 504 ± 58 0.026 ± 0.003
500 190 ± 6 491 ± 29 0.025 ± 0.002
800 208 ± 6 560 ± 40 0.029 ± 0.002
Revaclear 350 153 ± 6 334 ± 24 0.025 ± 0.002
500 176 ± 8 422 ± 48 0.031 ± 0.004
800 186 ± 8 415 ± 35 0.030 ± 0.003
Revaclear MAX 350 167 ± 7 434 ± 48 0.026 ± 0.003
500 179 ± 6 435 ± 31 0.026 ± 0.002
800 200 ± 6 520 ± 39 0.030 ± 0.002
a

Clearances based on rate of appearance in dialysate.

b

Calculated using the plasma water flow rate (QPWi).

Data are presented as mean ± SEM for n = 12 observations. Statistical differences are described in the text.