Table 3.
Phosphate
Dialyzer | Dialysate flow rate (mL/min) | Clearance (mL/min)a | KoA (mL/min)b | Ko (cm/min)b |
---|---|---|---|---|
Optiflux F160NR | 350 | 168 ± 6 | 434 ± 34 | 0.028 ± 0.002 |
500 | 178 ± 9 | 450 ± 62 | 0.029 ± 0.004 | |
800 | 194 ± 8 | 458 ± 33 | 0.029 ± 0.002 | |
Optiflux F200NR | 350 | 178 ± 8 | 504 ± 58 | 0.026 ± 0.003 |
500 | 190 ± 6 | 491 ± 29 | 0.025 ± 0.002 | |
800 | 208 ± 6 | 560 ± 40 | 0.029 ± 0.002 | |
Revaclear | 350 | 153 ± 6 | 334 ± 24 | 0.025 ± 0.002 |
500 | 176 ± 8 | 422 ± 48 | 0.031 ± 0.004 | |
800 | 186 ± 8 | 415 ± 35 | 0.030 ± 0.003 | |
Revaclear MAX | 350 | 167 ± 7 | 434 ± 48 | 0.026 ± 0.003 |
500 | 179 ± 6 | 435 ± 31 | 0.026 ± 0.002 | |
800 | 200 ± 6 | 520 ± 39 | 0.030 ± 0.002 |
Clearances based on rate of appearance in dialysate.
Calculated using the plasma water flow rate (QPWi).
Data are presented as mean ± SEM for n = 12 observations. Statistical differences are described in the text.