Skip to main content
. 2010 Oct 4;398(7):3115–3125. doi: 10.1007/s00216-010-4231-5

Fig. 2.

Fig. 2

The Zn isotope data (δ66Zn values) determined for London Zn–DS mixtures vs. time of data acquisition for two measurement sessions that had duration of a 3 h and b 23 h. In b, the measured δ66Zn values (filled symbols) are corrected for both the absolute offset and the drift in mass bias behavior (open symbols). This improves the precision of the dataset to ±0.04‰ (2 SD, open symbols)