Table 1.
List of parameters and variables.
Label | Definition (typical value) |
---|---|
Constraints | |
Tmax | maximum allowable temperature (5.25 °C) |
Tz | desired peak temperature in a treatment zone (5 °C) |
tp | pulse length (period = tp/dc) (50 ms) |
M | number of treatment zones (4-6) |
m | number of focal spots per treatment zone (8) |
tq | time of mechanical motion (30 s); q = tq/tp |
N | total number of pulses per focal spot (100) |
| |
Variables & Symbols | |
| |
T | temperature |
dc | device duty cycle |
DCz | local ‘treatment zone’ duty cycle |
DCN | net duty cycle |
nz | pulses per focal spot in a single pass |
n | pulse number, from 0 to N |
theat = mnztp/dc | time spent treating (heating) a single treatment zone |
tM = M (mnztp/dc + tq) | time required to treat M treatment zones |
tcool = tM – theat | time between treatments of a single treatment zone |
Δ | increase in Tmax over Tz: |