Table 1.

IEC and FDA guidelines on SAR and heating in human MRI studies

Limit Whole-Body
Average
Heat
Average
Head, Trunk
Local SAR
Extremities
Local
IEC (6-minute average)
   Normal (all patients) 2 W/kg (0.5°C) 3.2 W/kg 10 W/kg 20 W/kg
   First level (supervised) 4 W/kg (1°C) 3.2 W/kg 10 W/kg 20 W/kg
   Second level (IRB approval) 4 W/kg (>1°C) >3.2 W/kg >10 W/kg >20 W/kg
   Localized heating limit 39°C in 10 g 38°C in 10 g 40°C in 10 g
FDA 4 W/kg for
  15 min
3 W/kg for
10 min
8 W/kg in 1g
for 10 min
12 W/kg in 1g
for 5 min

Note: FDA = US Food and Drug Administration; IEC = International Electrotechnical Commission; IRB = institutional review board; MRI = magnetic resonance imaging; SAR = specific absorption rate.

Source: FDA [2], European Committee for Electrotechnical Standardization [3].