TABLE 2.
COMPARISON OF INTRA-DAY AND INTERDAY PRECISION AND ACCURACY
| Amount | HPLC method |
UV spectrophotometric |
||||
|---|---|---|---|---|---|---|
| Mean area | SD | RSD% | Mean area | SD | RSD% | |
| Within day precision | ||||||
| 20 | 933.55 | 5.92 | 0.634 | 0.205 | 0.001 | 0.357 |
| 40 | 1576.12 | 1.09 | 0.069 | 0.229 | 0.001 | 0.069 |
| 60 | 2350.64 | 9.47 | 0.398 | 0.288 | 0.001 | 0.173 |
| (Acceptance Criteria RSD<2) | ||||||
| Between day precision | ||||||
| 20 | 923.843 | 5.591 | 0.601 | 0.204 | 0.001 | 0.283 |
| 40 | 1585.137 | 11.175 | 0.704 | 0.231 | 0.003 | 1.5 |
| 60 | 2355.138 | 12.788 | 0.542 | 0.286 | 0.004 | 1.415 |
| (Acceptance Criteria RSD<3) | ||||||
Comparision of intra-day and inter-day precision and accuracy for the determination of zonisamide 100 mg by HPLC and UV-spectrophotometric methods