. 2006 Apr-Jun;31(2):67–71. doi: 10.4103/0971-6203.26690

Table 3.

Beam attenuation due to PMMA base plate (12 mm) without air gap and carbon fiber base plate

Gantry angle (degree) Path length (mm) PMMA PMMA base plate on the phantom % attenuation Path length (mm) Carbon fibre Carbon fiber plate on the phantom % attenuation
0 12.00 4.85 8.0 3.68
10 12.185 5.19 8.12 3.70
20 12.77 5.28 8.51 3.93
30 13.856 5.78 9.23 4.23
40 15.66 6.28 10.44 5.00
50 18.67 7.81 12.44 6.22
60 24.00 10.55 16.0 7.98
70 35.00 10.31 0.0 0.00
80 0.0 0.00 0.0 0.00
90 0.0 0.00 0.0 0.00

It is to mention that half of the beam is passing through the PMMA base plate at gantry angle 70 degree