Skip to main content
. 2006 Apr-Jun;31(2):67–71. doi: 10.4103/0971-6203.26690

Table 4.

TPS-calculated MUs for 100cGy dose at isocenter in phantom vs. measured dose

Gantry Angle TPS Cale. MU contour with base late Measured dose cGy TPS Cale. MU Contour Phantom alone Measured dose cGy % diff in dose delivery
0 123 98.406 119 95.13 3.33
10 124 98.50 119 94.31 4.25
20 126 98.859 120 93.22 5.70
30 133 98.95 125 92.95 6.06
40 139 98.45 132 92.41 6.13
50 154 98.72 143 90.77 8.05
60 154 98.81 154 98.95 −0.14
70 161 98.99 161 98.86 0.13