. 2010 Nov 1;4(6):1495–1503. doi: 10.1177/193229681000400625

Table 4.

Subject Errors

Error Subjects with this error, na Failure—no A1C result, nb FTF rate associated with error (n = 27)c
+DVD –DVD Total +DVD –DVD Total +DVD –DVD Total
Did not insert blood collector into shaker properly 9 17 26 4 11 15 15% 41% 56%
Did not fill blood collector properly 2 12 14 1 7 8 4% 26% 30%
Shook shaker for <5 s 12 8 20 2 3 5 7% 11% 19%
Did not shake the shaker at all 1 1 2 0 0 0 0 0 0
Removed shaker base too early 0 1 1 0 1 1 0 4% 4%
a

Subjects may have experienced more than one error during the initial self-test.

b

A total of 27 subjects experienced 29 errors that led to failure.

c

Calculated as (100 × number of failures associated with the error)/total number of subjects that experienced a failure.