TABLE 4.

Sensitivity and specificity of commercial IgG ELISAs compared to an anti-PT IgG in-house ELISAa

ELISA Sensitivity (%) Specificity (%)
Euroimmun Pertussis IgG 91/91 90/90
Virion\Serion PT IgG 89/78 70/85
Virotech PT IgG 84/80 87/93
Dr.Merk Pertussis IgG 100 51
IBL Pertussis IgG 100 52
MP Biomedical Pertussis IgG 100 51
Novagnost Pertussis IgG 94 55
Novatec Pertussis IgG 95 54
Trinity Biotech Pertussis EIA 91 59
Virion\Serion Pertussis IgG 86 56
Virotech Pertussis IgG 87 81
a

Indeterminate results according to the kit's insert were interpreted as positive. Sensitivity and specificity of kits measuring anti-PT were also estimated when the in-house ELISA cutoff (≥40 IU/ml) was used (second number where two numbers are given).