Skip to main content
. 2010 Nov 30;9:346. doi: 10.1186/1475-2875-9-346

Table 2.

Cell counts and data loss with a Cell-Dyn 3000 instrument.

Cell type Mean cell count in CBC result Mean number of cells analysed Mean number pixels on screen* Information (cell count) lost in scatter-plot display (%)**
WBCs 8675/μL 9100 482 94.4
Granulocytes 3834/μL 4174 348 90.9
Lymphocytes 2992/μL 3068 22 99.3
Monocytes 1334/μL 1379 58 95.7
Eosinophils 349/μL 345 43 87.7

Data shown correspond to the mean values for 153 CBC results from malaria positive patients, obtained with a CD 3000 instrument in Lambaréné, Gabon. *Coloured pixels indicate the events that represent analysed cells in the side-scatter/depolarized side-scatter plot (lobularity/granularity) on-screen (Figure 1). ** % data lost = 1- (Mean number pixels on screen/Mean cell count in CBC result) × 100. Screen images were analysed by taking a screenshot and counting the pixels using ImageJ image processing software.