Skip to main content
. 2010 Oct 22;41(1):55–75. doi: 10.1007/s00247-010-1858-5

Table 3.

Guidelines for tube voltage (kVp) and effective tube current (mA) for 16-slice CT scanners [6]

Weight CTDIvol kVp* GE Philips Siemens Toshiba
16 × 1.25 mm 16 × 0.75 mm 16 × 0.75 mm 16 × 1 mm,M
4–9 kg 2.1 100 (90) 80 120 80 50
10–19 kg 3.0 100 (90) 110 170 115 70
20–29 kg 3.8 100 (90) 140 215 145 90
30–39 kg 4.1 100 (90) 150 235 155 95
40–49 kg 4.9 120 100 125 120 95
50–64 kg 5.9 120 120 150 140 115

*tube voltage (kVp) for Philips scanners between brackets