. 2010 Nov 17;49(1):118–124. doi: 10.1128/JCM.00685-10

TABLE 2.

Performance of HIV-1 viral load assays according to HIV subtype

Parameter No. (%) of patients with indicated result based on:
m2000rt TaqMan 48 NucliSens Easy Q v1.2 Versant 3.0
No. of overall results obtained (n = 369 patients) 371 (100) 357 (97) 338 (92) 276 (75)
TND or no amplificationa 74 (20) 69 (19) 23 (7) NA
    Subtype B 12 (16) 10 (14) 13 (57)
    Non-subtype B 62 (84) 59 (86) 10 (43)
<LDL log10 copies/ml 28 (7.5) 24 (7) 74 (22) 109 (39)
    Subtype B 7 (25) 7 (29) 16 (22) 20 (18)
    Non-subtype B 21 (75) 17 (71) 58 (88) 89 (82)
False-negative resultb (n = 38 samples) 8 (2) 9 (2.5) 20 (6) 5 (2)
    Subtype B 4 (50) 6 (67) 7 (35) 0
    Non-subtype B 4 (50) 3 (33) 13 (65) 5 (100)
Underquantified result ≥1.0 log10 copies/mlc (n = 50 samples) 5 (1.3) 7 (2) 15 (4.4) 26 (9.4)
    Subtype B 2 (25) 3 (43) 4 (27) 5 (19)
    Non-subtype B 3 (75) 4 (57) 11 (73) 21 (81)
Underquantified result ≥0.5 and <1.0 log10 copies/mlc (n = 95 samples) 23 (9) 20 (6) 61 (18) 68 (25)
    Subtype B 11 (48) 13 (65) 26 (43) 26 (38)
    Non-subtype B 12 (52) 7 (35) 35 (57) 42 (62)
Falsely high (VL ≥1.0 log10 copies/ml [n = 4]) None 1 CRF02_AG 1 C 2 (1) (1 B and 1 AE)
Total discordant resultsd 13 (3.5) 17 (5) 36 (11) 33 (12)
a

All of the real-time assay results are reported as target not detected (TND) or no amplification (EQ) result if no HIV-1 RNA was detected. NA, not applicable.

b

Defined as a TND result for a real-time assay where all of the other assays detected HIV-1 RNA above the LDL or where an assay gave a result <LDL while all of the other assays gave an HIV-1 VL of > 2.0 log10 copies/ml higher. Three samples were false negative by more than one assay.

c

A total of 144 samples were underquantified by one or more assays as follows (number of samples underquantified in parentheses): m2000rt (8), CAP-CTM (8), EQ (40), bDNA (36), m2000rt + EQ (2), m2000rt + CAP-CTM (2), m2000rt + bDNA (10), CAP-CTM + EQ (7), CAP-CTM + bDNA (2), EQ + bDNA (22), CAP-CTM + EQ + bDNA (3), and m2000rt + EQ + bDNA (4).

d

Results include false negatives, falsely high results, and results underquantified by ≥1.0 log10 copies/ml. The total number of results for each assay was used to calculate the percentages.