Table 8.

Prosthetic aortic valve stenosis criteriaa

Parameter Normal Possible stenosis  Significant stenosis
Peak velocity (m/s)b <3 3–4 >4
Mean gradient (mmHg)b <20 20–35 >35
Doppler velocity index ≥0.30 0.29–0.25 <0.25
Effective orifice area (cm2) >1.2 1.2–0.8 <0.80
Contour of the jet velocity through the prosthetic valve Triangular, early peaking Triangular to intermediate Rounded, symmetrical contour
Acceleration time (ms) <80 80–100 >100

aIn conditions of normal or near normal stroke volume (50–70 mL).

bThese parameters are more affected by flow, including concomitant aortic regurgitation.