TABLE III.

Measured Crimped and Expanded Values Obtained with the Solid Tubular and the Laser Etched Stents

Stent Type

Solid Tubular Laser Etched
Crimped Diameter (mm) 1.85 ± 0.04 1.40 ± 0.03
Length (mm) 18.5 ± 0.20 17.3 ± 0.50
Expanded Diameter (mm) 3.96 ± 0.03 3.88 ± 0.08
Thickness (mm) 0.25 ± 0.02 0.25 ± 0.02
Length (mm) 17.87 ± 0.03 17.12 ± 0.74
Axial shortening (%) 3.41 ± 0.90 1.14 ± 1.60
Radial expansion ratio 2.14 ± 0.08 2.72 ± 0.11