The rms deviation from the mean and maximum deviation from the mean, comparing the raw fluorescent image intensity, the uncorrected probe measurement at 635 nm, and the model-derived fluorophore concentration for phantoms A to I at [PpIX] = 5 μg∕mL.

  rms Deviationfrom Mean (%) Maximum Deviationfrom Mean (%)
Fluorescence image intensity 56.5 150.2
Measured fluorescence Fx,mm = 635 nm) 52.3 94.9
Estimated [PpIX] 10.1 14.2