Table 2.

Interassay precision

SRR%

Serum Mean s.d. Range CV%
High TSI 446 51 331–596 11·5
Medium TSI 263 34 221–371 12·8
Low TSI 158 23 117–193 14·5
Negative 52 8 40–74 15·7

Thyroid-stimulating immunoglobulin (TSI)-containing sera of Graves' disease patients were diluted with negative control serum (negative) to yield percentage of specimen-to-reference ratio (SRR%) values near the cut-off (low), slightly higher that the cut-off (medium) and near the upper level of the TSI-Mc4 bioassay detection (high). Each sample was tested repeatedly 120 times by a single user over a period of 20 days. CV, coefficient of variation; s.d., standard deviation.