Table 2.
Interassay precision
| SRR% | ||||
|---|---|---|---|---|
| Serum | Mean | s.d. | Range | CV% |
| High TSI | 446 | 51 | 331–596 | 11·5 |
| Medium TSI | 263 | 34 | 221–371 | 12·8 |
| Low TSI | 158 | 23 | 117–193 | 14·5 |
| Negative | 52 | 8 | 40–74 | 15·7 |
Thyroid-stimulating immunoglobulin (TSI)-containing sera of Graves' disease patients were diluted with negative control serum (negative) to yield percentage of specimen-to-reference ratio (SRR%) values near the cut-off (low), slightly higher that the cut-off (medium) and near the upper level of the TSI-Mc4 bioassay detection (high). Each sample was tested repeatedly 120 times by a single user over a period of 20 days. CV, coefficient of variation; s.d., standard deviation.