Table 3.

Variations in the depth to the distal tumor volume surface (D2DT) correlated with variations in position of the distal prescription iso-dose. (The data are for the right lateral beam only). Statistically significant correlations with the D2DT and the femur setup angle are highlighted in bold.

Case Plan CT
In-room CT
Proton range [mg/mm2] SOBP mod [mg/mm2] D2DT [mm] D2DT range [mm] D2DT mean ± standard deviation [mm] Deviation in the distal iso-dose [mm] Correlation p-value
D2DT Femur angle
1 252 74 220 ± 1 214–228 221 ± 3 −0.4 ± 3.1 0.001 0.107
2 295 88 277 ± 2 274–290 281 ± 4 2.1 ± 2.9 0.169 0.022
3 248 65 226 ± 1 223–231 228 ± 2 2.3 ± 2.3 0.052 0.001
4 272 80 241 ± 1 237–254 244 ± 3 2.5 ± 2.0 0.023 0.199
5 224 64 202 ± 1 200–208 204 ± 2 3.0 ± 2.4 0.049 0.002
6 232 80 206 ± 1 203–212 206 ± 2 2.7 ± 2.5 0.009 0.309
7 218 92 192 ± 1 187–195 191 ± 2 −0.4 ± 2.1 0.021 0.063
8 236 86 200 ± 1 200–206 204 ± 1 −3.4 ± 1.5 0.029 0.372
9 263 75 230 ± 1 228–236 232 ± 2 −0.7 ± 1.9 0.001 0.857
10 233 76 207 ± 1 197–208 204 ± 2 2.8 ± 2.4 0.023 0.308

Abbreviations: CT = computed tomography scan, SOBP = spread-out Bragg peak, D2DT = depth to distal tumor volume surface