Skip to main content
. 2010 Dec 10;37(2):233–240. doi: 10.1007/s00134-010-2098-8

Table 2.

Mean bias and limits of agreements (95% confidence interval)

Bias % (95%CI) Percentage error % (95%CI)
All measurements (58 patients/401 measurements), without correction
 G3 − iCO −2.6 (−4.1 to −1.2*,†) 29.2 (25.2–34.2)
 G2 − iCO −12.4 (−14.0 to −10.8) 32.5 (26.3–36.7)
 CCO − iCO 8.2 (7.0–9.5)* 25.4 (21.3–29.5)
All measurements (58 patients/401 measurements), with correction [23, 24]
 G3 − iCO 0.2 (−3.7 to 4.2)*,† 30.4 (23.6–37.2)
 G2 − iCO −10.3 (−15.4 to −5.3) 28.6 (20–37.2)
 CCO − iCO 9.5 (5.8–13.1)* 28.0 (21.8–34.2)
First measurement only (58 patients/58 measurements)
 G3 − iCO −2.6 (−6.4 to −1.1)*,† 29.2 (22.7–35.7)
 G2 − iCO −12.4 (−16.6 to −8.2) 32.8 (27.5–38.1)
 CCO − iCO 8.2 (5.0–11.5)* 25.6 (19.9–31.3)

iCO cardiac output measured by bolus thermodilution; CO G2 cardiac output measured by second-generation FloTrac; CO G3 cardiac output measured by third-generation FloTrac; CCO cardiac output measured by semi-continuous thermodilution

p < 0.05 versus G2, p < 0.05 versus CCO