Published in final edited form as: Phys Med Biol. 2010 Feb 4;55(5):1279–1293. doi: 10.1088/0031-9155/55/5/002

Table 2.

An assessment of the effect of inter-field dose contamination when measuring Sc, p in the field geometry in figure 1(a). The ‘combined geometry’ is shown in figure 1(a). The ‘separate geometry’ refers to the irradiation of one piece of film per field

Field size (mm) EBT (combined geometry) EBT (separate geometry) Percent error
30 × 30 0.901 ± 0.018 0.892 ± 0.018 1.0%
20 × 20 0.899 ± 0.018 0.886 ± 0.018 1.5%
10 × 10 0.802 ± 0.016 0.795 ± 0.016 0.8%
5 × 5 0.642 ± 0.013 0.636 ± 0.013 0.7%