Table 7.

CT dose and induced σNCT for a range of imaging protocols (Baxter et al 2007).

Imaging
protocol
Phantom kVp mAs Slice
(mm)
Dose/image
(cGy)
No of image
averages
ΔNCT
Volumetric 16 cm diam. 120 200 2 3.38 ± 0.14 16 0.105 ± 0.001
(CTDI) (center) 5 3.4 ± 0.1 16 0.105 ± 0.001
10 3.4 ± 0.1 16 0.107 ± 0.001
16 cm diam. 140 200 2 3.5 ± 0.1 16 0.109 ± 0.001
(edge) 5 3.5 ± 0.1 16 0.110 ± 0.001
10 3.6 ± 0.1 16 0.111 ± 0.001
16 cm diam. 140 200 2 4.4 ± 0.2 16 0.136 ± 0.002
(center) 5 4.4 ± 0.2 16 0.136 ± 0.002
10 4.5 ± 0.2 16 0.138 ± 0.002
16 cm diam. 140 200 2 4.5 ± 0.2 16 0.139 ± 0.002
(edge) 5 4.5 ± 0.2 16 0.139 ± 0.002
10 4.6 ± 0.2 16 0.141 ± 0.002
Single slice 16 cm diam. 120 200 2 0.70 ± 0.05 32 0.0430 ± 0.0002
(PD) (center) 5 0.91 ± 0.06 32 0.0560 ± 0.0004
10 1.07 ± 0.07 32 0.066 ± 0.001
16 cm diam. 120 200 2 1.6 ± 0.1 32 0.099 ± 0.001
(edge) 5 1.7 ± 0.1 32 0.107 ± 0.001
10 2.1 ± 0.2 32 0.133 ± 0.002
16 cm diam. 140 200 2 0.83 ± 0.06 32 0.0510 ± 0.0003
(center) 5 1.07 ± 0.07 32 0.066 ± 0.001
10 1.23 ± 0.09 32 0.076 ± 0.001
16 cm diam. 140 200 2 1.5 ± 0.1 32 0.095 ± 0.001
(edge) 5 2.0 ± 0.1 32 0.125 ± 0.002
10 2.1 ± 0.2 32 0.130 ± 0.002
Calibration Calibration 140 10 2.3 ± 0.2 16 0.0720 ± 0.0001
(PD) gel vials