Table II.

Statistical Precision of 31 Measurements of 10 μm Latex Beads at Various Concentrations

Average particle count (n = 31) Standard deviation CV (%)
106 9.5 8.9
1,076 31 2.9
9,755 86 0.9
10,005 85 0.9
10,500 99 1.0
49,444 221 0.5
98,559 234 0.2