Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2008 Dec 24;877(3):285–290. doi: 10.1016/j.jchromb.2008.12.024

Table 5.

Stability of INH stock and plasma samples.

storage conditions INH sample(ng/mL) % remaining % RSD
2 freeze/thaw cycles Low (180) in plasma 95.1 1.2
High (9000) in plasma 95.0 2.0
freezing for 27 days Low (180) in plasma 97.1 0.6
High (9000) in plasma 92.1 9.0
benchtop 1 hr Low (180) in plasma 100.4 3.4
High (9000) in plasma 100 0.8
benchtop 3 hr Low (180) in plasma 83.4 3.1
High (9000) in plasma 91.2 2.7
benchtop 20 hr Low (180) in plasma 63.2 6.6
High (9000) in plasma 61.7 3.0
autosampler (72 hr) Low (180) in extract 93.4 6.2
High (9000) in extract 89.7 2.3
freezing for 6 months benchtop 20 hr INH Stock solution 101 2.1
INH Stock solution 99.0 5.5