Table 3.

Intra-day and inter-day precision (%RSD) and accuracy (% dev) for analysis of artemether and dihydroartemisinin in human plasma.

Intra-day
Inter-day
Nominal, ng/mL 2 6 80 170 2 6 80 170
ARM mean, ng/mL 1.8-2.3 5.2-6.4 70.5-86.8 160-182 1.96 5.98 76.8 170
SD 0.1-0.2 0.5-0.6 4.0-8.7 4.7-14.2 0.22 0.63 7.7 13
RSD,% 5.2-8.0 2.9-12 2.0-7.7 2.9-7.9 11 10 10 7.5
%dev −11-13 −13-7.1 −12-9.1 −5.8-6.8 −2.1 −0.3 −4.0 0.1
n 5 5(4) 5 5 25 24 25 25
DHA mean, ng/mL 1.9-2.1 5.2-5.7 72.5-79.9 152-165 2.00 5.42 74.9 158.2
SD 0.1-0.1 0.1-0.3 1.8-4.6 4.0-13 0.11 0.27 3.8 8.4
RSD,% 2.1-5.8 1.8-5.7 1.9-6.0 2.5-8.2 5.3 5.0 5.1 5.3
%dev −3.3-5.6 −13- −4.8 −9.3- −0.1 −10- −3.2 0.1 −9.7 −6.4 −6.9
n 5 5 5 5 25 25 25 25