Technical Pearls for Amplatzer Vascular Plug Embolization

Issue Pearl
Oversizing Oversize the device by 15–30% (30–50% for the single disc device) relative to the vessel to prevent distal migration.
Positioning Two platinum bands serve as radiopaque markers.
Deployment A simple unscrewing mechanism allows for precise deployment of the device.