Table 2.
Sampler pairs | Sample number | Correlation coefficient (P value) | Average difference between the groups (mg m−3) (P value)a | Correction factorsb |
ACCU-CAP™/ACCU-CAP™ | 29 | 0.26 (0.264)c | 6.66 (0.324)c | 0.16 ± 0.50c |
ACCU-CAP™/Button | 30 | 0.94 (<0.001) | 0.10 (0.885) | 0.91 ± 0.46 |
ACCU-CAP™/CIP10-I | 30 | 0.87 (<0.001) | 1.08 (0.267) | 0.60 ± 1.52 |
ACCU-CAP™/GSP | 33 | 0.95 (<0.001) | 0.98 (0.709) | 0.77 ± 0.75 |
ACCU-CAP™/IOM | 32 | 0.46 (0.008) | 0.55 (0.270) | 0.64 ± 0.38 |
Button/Button | 30 | 0.96 (<0.001) | 0.03 (0.933) | 0.97 ± 0.34 |
Button/CIP10-I | 32 | 0.22 (0.219) | 0.33 (0.529) | 0.82 ± 1.18 |
Button/GSP | 30 | 0.32 (0.087) | 0.29 (0.404) | 0.79 ± 0.57 |
Button/IOM | 24 | 0.80 (<0.001) | 0.05 (0.913) | 0.95 ± 0.52 |
CIP10-I/CIP10-I | 24 | 0.94 (<0.001) | 0.27 (0.714) | 1.13 ± 1.02 |
CIP10-I/GSP | 30 | 0.70 (<0.001) | 0.11 (0.871) | 1.08 ± 0.35 |
CIP10-I/IOM | 28 | 0.99 (<0.001) | 0.22 (0.924) | 1.08 ± 0.39 |
GSP/GSP | 30 | 0.91 (<0.001) | 0.17 (0.734) | 0.88 ± 0.26 |
GSP/IOM | 31 | 0.91 (<0.001) | 0.03 (0.982) | 0.99 ± 0.39 |
IOM/IOM | 31 | 0.74 (<0.001) | 0.75 (0.573) | 0.74 ± 0.74 |
Average difference between the groups was calculated using absolute value difference between the sampler groups.
Correction factors is applied to the second sampler to make average mass concentration same as the first sampler in sampler pairs; ± indicates 95% confidence interval value.
Correlation coefficient becomes 0.83 (P < 0.001), the average difference becomes 0.01 mg m−3 (P = 0.98), and the correction factor becomes 1.02 ± 0.50 with the removal of a single outlier.